Buckled Upper Pole Breast Style 410 Implant
Date: Tue, 13 Jun 2000 00:49:27 -0500
From: Rogene
saxony01@netscape.netTo:
delphine1939@videotron.caBuckled Upper Pole Breast Style 410 Implant
Aesth. Plast. Surg. 23:279-281 (1999)
© 1999 by Springer-Verlag New York, Inc.
Buckled Upper Pole Breast Style 410 Implant
Presenting as a Manifestation of Capsular
Contraction
Darryl James Hodgkinson
----------------------------------------------------
Abstract
Recent experience with the cohesive gel implant has shown an interesting variation of capsular contracture. Instead of spherical contraction, the [space] capsule formation around the 410 implant seems to cause a buckling of the superior pole of the implant. This buckling has been observed in four implants and presented clinically in 2 of 50 patients with a superior pole mass. The capsular contraction around the implant and the buckling are probably camouflaged by a subpectoral implantation and this presentation should be recognized by the surgeon, especially when considering placement of the cohesive gel implant in a subglandular position. Histological sections of a capsule around the implant show fragmented silicone, with a cellular reaction around the fragmented silicone.
Bryan